Model: 8031 Specification: l.Assay methods: End Point, Kinetic, Fix time etc. 2.Principle: Photoelectric colorimetry 3.Light Source: Halogen lamp 12V/20W 4.Photometry range: 0-3.2Abs 5.Resolution: 0.0001Abs 6.Wavelength: 10 wavelengths optional (340nm-800nm) 7.Throughput: 300tests/hour 8.Reagent tray: 59 reagent positions, 1 detergent position 9.Sample tray: 71 sample positions, including detergent, standard, QC, STAT positions 10.Reaction tray: 120 reaction cuvettes 11.Sample volume: 2-100u1, with 0.1u1 increment Reagent volume: R1:10-500uL, R2: 10-500uL,with 0.5uL increment Minimum reaction volume: 180uL Maximum reaction time: 10 minutes Water Consumption: 6L/hour under working status 12.Clean unit: 8-step auto-washing system with detergent and 37 C water 13.Calibration: Calibration reset, select best test point by reaction curve, no need second calibration Line/non-line;multi-standards assay 14.Control rules: Westgard multi-rule, Cumulative sum check, Twin plot 15.3 level controls for each item, analyzing and printing QC analysis diagram